Innovating science for over 125 years

Forced Convection Incubator

IN602C/612C/802C/812C

Description

Standard low temperature forced convection incubators

Operation and functions

  • Large capacity
  • Forced circulation system enables high accuracy temperature control and even temperature distribution
  • All models come with either an observation window (W) for improved visibility or solid door
  • Inner glass door keeps  temperature stable during sample observation
  • Employment of a large dual-glass door and inner door, forms a triplex glass doors for better heat insulation result (for models with observation window)
  • Equipped with various functions such as self-diagnostic (temperature sensor error, heater disconnection, SSR short-circuit, automatic overheat prevention), calibration off-set, clock, timer, overheat prevention device and key lock
  • Manual and program defrost operation are both available
  • After power failure, unit automatically restarts operation at setting before power failure
ModelIN602C
IN612C
IN802C
IN812C
SystemForced convection
Operating temperature range-10°C~+50°C
Temperature adjustment accuracy±0.3°C (when refrigerator in continuous operation) / ±1.0°C (when refrigerator in cycle operation)
Temp. distribution accuracy±1.0°C (when refrigerator in continuous operation at 37°C)
Maximum temp. reaching time20~50°C ~20min20~50°C ~30min
Minimum temp. reaching time20~-10°C ~ 45min
20~-10°C ~55 min for IN602CW / 612CW
20~-10°C ~65min
InteriorStainless steel
ExteriorChrome free electronic galvanized plated steel plate chemical proof baking finish
Heat insulation materialStyrene foam
RefrigeratorAir-cooled fully closed compressor 275WAir-cooled fully closed compressor 375W
Refrigerator mediumR134A
Defrosting mechanismManual ON / Auto OFF, Timer operation, Cycle operation
Blower fanAxial fan
HeaterIron-chrome wire heater: 550WIron-chrome wire heater: 750W
SensorPlatinum resistance temperature detector: Pt100Ω (temperature controller), K-thermocouple (overheat prevention device)
Cable portI.D.: 32 mm (right side)
Temperature controlPID control
Temperature settingDigital setting with ▲▼ keys
Temperature displayMeasured temperature: 4-digit orange LED digital display + VFD fluorescent display
Timer / timer resolution0 min. ~ 999 hrs. 59 min. / 1 min.
Operation functionFixed temperature, auto stop, auto start, program (up to 32 steps, repeat operation)
Additional functionTimer function (accumulated time to 49,999 hrs), calibration off-set function, clock display
Safety deviceRefrigerator overload relay, independent overheat prevention device, over current leakage breaker, self diagnostic functions (temperature sensor error, heater disconnection, SSR short-circuit, automatic overheat prevention), abnormal buzzer alarm
Internal dimensions W x D x H600 x 477 x 500 mm600 x 477 x 1000 mm
External dimensions W x D x H710 x 645 x 915 mm710 x 645 x 1630 mm
Internal capacity143L286L
Shelf plate load15kg / pc.
Shelf rest step number / pitch13 steps / 30mm23 steps / 30mm
Power supply (50/60Hz)AC115 9A
AC220V 5A
AC115 10A
AC220V 6A
Weight~89kg~ 115kg
Shelf / Shelf brackets3 pcs. / 6 pcs.5 pcs. / 10 pcs.
Optional itemsStand, additional shelf, cable port (ø30/50mm), recorder, warning light combination (stand-by / operation / error), observation window, external communication function (RS485), temperature output terminal (4-20mA), external alarm output terminal, time up output terminal

in-detail

Click on the links below to download the associated PDFs.


Incubator Overview

IN Series Catalog

To download Operational Manual, visit our Technical Support Center