Evaporated water amount
Max. 1,300 mL/h
Temp. adjustment range
40 to 220deg.C
Sample flow
Variable up to 26 mL/min.
Spray nozzle (selectable)
For liquid/gas

Easily grinds sample into fine powder with the spray drying system

GB210: This model supports water soluble and organic solvents* (*When organic solvent is used, a GAS410 type organic solvent recovery unit will be necessary). The unique attribute of this spray dryer is its ability to use different glassware sets allowing for a granulation mode.

Specifications
Performance Water Evaporation Rate Max. approx. 1,300 ml/h
Temperature Control Range 40˚C - 220˚C at inlet
Temperature Control Accuracy ± 1°C at inlet
Dry Air Flow Rate Max. 0.7 m³/min
Air Spray Pressure Control Range 0 - 249 k Pa (0-3 kg/cm²)
Flow Rate for Sample Liquid Feed Pump Max. 26 ml/min
Structure Temperature Controller PID digital tempertaure controller
Temperature Sensor K thermocouple
Stainless Tube Heater 2.88 kW at 240V
Sample Liquid Feeding Pump Quantitative peristaltic pump
Solvent Recovering Capability Organic solvent recovery unit GAS-410 must be used
Spray Line Cleaning Needle inside the nozzle to clean the mesh automatically 
Safety Devices Self-diagnostic functions (e.g. temperature aberration); Sample feed reversal
Standard Air Spray Pressure Gauge Bourdon tube: 294k Pa (3 kg/cm²)
External Dimensions (W x D x H) 760 x 420 x 1350 mm or                                   30 x 16.5 x 53.2 in
Weight 110 kg or 243 lbs
Power Supply  AC 200V - 240V, single-phase 18 A
Included Accessories Sample liquid tube Silicone tube - 2 pcs
Safety Cover Yes
Air hose 1 pc
Exhaust Duct 1 pc
Required Accessories Mini Spray Attachment GF - 300 or GF - 200
Optional Accessories Organic Solvent Recovery Unit GAS - 410
Inlet/outlet Temperature Recorder 212792 - factory installed
Fine Powder Recovery Cyclone 212780
Viton/Tiron Feeding Tube W0180001
Nozzle 1A, 1, 2A, 2, 3
Operational Requirements Compressed Air 28 l/min or 3 kgf/cm²; not required when using GAS-410
Type of Gas N₂ gas (99.99% purity, medical grade) is required when using GAS-410


CONTACT USToll free 1-800-2-YAMATO/ E-mail: customerservice@yamato-usa.com