Inovando a ciência há mais de 130 anos

Solicitar um Formulário de Orçamento

DNE410 - DNE411 Energy Saving Forced Convection Ovens

Phased out:
Yamato DNE410
Upgrade to our new replacement product:
Yamato DNE411

Common Specifications:

  • Programável
  • Internal capacity: 90L
  • Temperature control accuracy: ± 0.5°C at 210°C
  • Temperature distribution precision: ± 2.0°C at 210°C
  • Safety devices: Self-diagnostic functions (detection for sensor error, heater disconnect, triac short, faulty main relay, overheat prevention activation), keypad lock function, independent overheat prevention device, surge protective breaker
  • Internal dimensions: 450×450×450 mm (W×D×H)
  • External dimension: 580×646×860 mm (W×D×H)
  • Power source: 220V
  • Feito no Japão

Old vs. New

Upgraded Product Specifications

ModeloPhased-out Model DNE410Replacement Model DNE411
Faixa de temperaturaRT + 10 ° C a 210 ° CRT + 20 ° C a 210 ° C
Temperature control deviceModel IVCRModel V for improved temperature control and added functions
Função de operaçãoFixed temperature, Program operation (max. 32 steps, repeat operation), Auto start, Auto stopFixed temperature, Program operation (max.99 steps, up to 99 patterns repeat operation), Duration / time select timer operation function (fixed temperature operation, auto-start, auto-stop, quick auto stop, program operation auto start)
Funções adicionaisAccumulated power-on / operation time (up to 65535 hours), clock timer (24-hr), calibration offsetAccumulated power-on / operation time (up to 65535 hours), clock timer (24-hr), calibration offset, power failure recovery mode select, power consumption/CO2 emissions/heater output monitoring, user settings data save & restore
Itens opcionaisStand, stacking kit, punched stainless shelf, 25 mm and 50 mm diameter cable port, observation window, temperature output terminal (4-20mA), remote communications adapter kit, external alarm output terminal, time-up output terminalStand, stacking kit, punched stainless shelf, 25 mm and 50 mm diameter cable port, observation window, temperature output terminal (4-20mA), remote communications adapter kit, external alarm output terminal, time-up output terminal, operation signal output terminal, event output signal terminal, sheath sensor (k-thermocouple), silicon plug (for ports)
Other enhancements of replacement modelDoor hinges changed to adjustable butterfly hinge to eliminate door seal gaps

Safety enhancement: moved sensor closer to heater to lessen overheat errors

For complete technical specifications of the phased-out model, contact Serviço ao cliente.