Yamato DNF411


Common Specifications:
- Programável
- Internal capacity: 90L
- Safety devices: Self-diagnostic functions (detection for sensor error, heater disconnect, triac short, faulty main relay, overheat prevention activation), keypad lock function, independent overheat prevention device, surge protective breaker
- Internal dimensions: 450×450×450 mm (W×D×H)
- External dimension: 580×645×860 mm (W×D×H)
- Power source: 220V
- Feito no Japão
Old vs. New
Upgraded Product Specifications
Modelo | Phased-out Model DNF410 | Replacement Model DNF411 |
---|---|---|
Sistema | Convecção forçada | Forced and natural convection |
Air velocity control system | Variable 10 stages (800 ~ 1500 rpm) | Variable 10 stages (600~1500 rpm) + Air speed level [0] |
Controlled temperature range | RT+10°C ~ 260°C | RT +15°C ~ 260°C (forced convection RT +25°C ~ 120°C (natural convection) |
Precisão de controle de temperatura | ± 0.5°C at 260°C | ± 0.3°C at 260°C (forced convection) ± 0.3°C at 120°C (natural convection) |
Precisão da distribuição de temperatura | ± 2.5°C at 260°C | ± 2.5°C at 260°C (forced convection) ± 3°C at 120°C (natural convection) |
Temperature control device | Model IVCR | Model V for improved temperature control and added functions |
Capacity / Number of heaters | 1.25 kW x 1 | 0.6kW x 2 |
Função de operação | Fixed temperature, Program operation (max. 32 steps, repeat operation), Auto start, Auto stop | Fixed temperature, Program operation (max.99 steps, up to 99 patterns repeat operation), Duration / time select timer operation function (fixed temperature operation, auto-start, auto-stop, quick auto stop, program operation auto start) |
Funções adicionais | Accumulated power-on / operation time (up to 65535 hours), Clock timer (24-hr), Calibration offset | Variable wind speed function Power on and Operation Time Integrating Function (up to 65,535 hours); Calibration Offset; Monitor Display of Integrated Power Consumption, Total CO2 Emission, and Heater operating Output; Power Recovery Mode; Save and Access of Operator’s Setting Information |
Other enhancements of replacement model | Heater shape change for compatibility to a 2 way convection Exhaust damper modified for increased ventilation efficiency Door hinges changed to adjustable butterfly hinge to eliminate door seal gaps |